Troubleshooting
Description
Result
Possible Cause
Remedy
Message displayed
showing
±
30 minutes
left of battery operation
Low Battery
Connect to charger
If pump
won’t be
connected
to mains,
battery will
be depleted
Message will be displayed as
long as the internal battery
voltage is less than 7.1v.
Connect to Mains
End Battery
End Battery
Connect to charger
Infusion
ceases
Battery is depleted.
The pump cannot be used
before being connected to
mains.
Connect to Mains and
wait 2 minutes before
operating.
Message displayed in the
TPN programs, that the
parameters entered are
incorrect and cannot be
calculated by the pump
Wrong TPN Date
Check Data
Program
will not
start
operating
The set parameters are
impossible to perform.
Volume or time parameters
are incorrect
Check data and change
accordingly.
Pump is in lock-mode
Lock out
Program
parameters
can not be
changed
The program is set in
second-level locking, which
enables daily operation of
the set program, without the
ability to change parameters.
No action required.
A fatal error is detected
Error XXX
Send for Service
Infusion
ceases
The pump will advise the
user to send the pump for
service.
Restart the pump. If the
alarm does not stop
pack the pump
properly, in its original
packaging, and send it
for service.