VET-PRO
Operator’s Manual
36
Program Setting
F
Note:
Whenever you decide to set a new program, follow the
instructions at the bottom of the screen.
Note
: -If one of the values is changed during the programing mode,
the pump will not store the new value, unless the change is
confirmed by pressing the
START/OK
key.
-If the flush key is activated during data setting, the pump
will return to the original (previous) screen, once flushing
has been completed.
Rate 4 ml/h
1.
The pump will display the last set rate. Change
or press
START/OK
to confirm.
Press OK for Volume
Dose Vol: 100 ml
2.
The pump will display the last set Dose
Volume. Change or press
START/OK
to
confirm
Press OK to continue
Pause 04:00
3.
Adjust the time pause between Doses (up to
20:00 hours), or confirm the last setting by
pressing the
START/OK
key.
Press OK to continue
Volume 1000 ml
4.
The pump will display the last set Volume
(total volume to be infused). Use the numeric
keys to change or press
START/OK
to
confirm and start program operation.
Press OK to start
5.
During the intermittent program the display
will show continually the actual rate of
infusion.
Rate
5 ml/h
6.
During a pause between the doses, the display
will show the KVO screen.
Delay 4:00
KVO 5 ml/h