Down Occlusion Setting:
1.
Switch the pump on, by pressing
ON/OFF
key
2.
When the screen of Rate appears, press
info
key twice.
Volume Infused
Show Setting
3.
Use
ARROW
keys (
2?, 0?)
to scroll to "Adjust
Setting" option.
=> Adjust Setting
4. Choose "Adjust Setting" with
START/ OK
key.
Adjust setting:
=> Pressure Level
5.
The option selected is “pressure level”.
Buzzer Level
6. Confirm with
START/OK
key.
Pressure setting
High
10 psi
7. Choose the desired pressure level, using the
arrow keys (
2?, 0?).
Adjust ?? & press OK
8. Press the
START/OK
key to store the parameter in the pump’s memory.
F
Note:
The following pressure parameters can be set:
Low = 5 psi (0.3 bar)
Normal = 7.5 psi (0.5 bar)
High = 10 psi (0.7 bar)
F
Note:
The selected pressure setting will remain in the pump’s memory
only until turning off (and then return to the default - HIGH).
F
Note:
In the event a down occlusion situation occurs, the patient is
protected by an anti-bolus feature, which prevents bolus delivery
to the patient when the downstream occlusion is released
.