VET-PRO
Operator’s Manual
43
Bag Vol: 500ml
Volume Left 500
Infused 0
3.
P
C
A
Resume - Press OK
Pressing the
START/OK
key
will display the
protocols list. Pressing the
STOP/NO
key will
display again PCA main menu.
A:MORPHINE
B: METHADONE
4.
Use the arrow keys to scroll through different
protocols and select with the
START/OK
key.
C: PETHIDINE
Rate 50 ml/h
BOLUS 5 ML
LockTime 60 Min
5.
Pump will display the protocol accumulation
screen. If Epidural mode was set to OFF,
pressing
START/OK
key will start program
operation.
P
A
4 H Limit 100 ml
Air sensor alarm
ON
Adjust ?? & press OK
6.
If epidural mode is On, pump will display Air
sensor alarm message. User can choose to set
air sensor to OFF (use arrow keys to change air
sensor status and press
START/OK
key for
approval). Press the
START/OK
key to start
operation.
MD Bolus
P
C
A
Flush
=>
MD Bolus
History
P
C
A
MD Bolus
Volume 2 ml
Press OK to start
Selecting the
MD Bolus
, from PCA menu,
will
display the MD Bolus setting screen. If no key
is pressed within 30 sec. the pump will go back
to the code screen.
Use the numeric keys to set the Bolus Volume,
up to 25.5 ml.
When the Bolus is completed the pump will
display the Resume/Menu screen. Press the
STOP/NO
key to go back to the menu code
screen or the
START/OK
key to display the
program accumulation screen.