VET-PRO
Operator’s Manual
52
Choose "Adjust Setting" with
START/OK
key.
Buzzer level
=> More …
Use
ARROW
keys to scroll to "More" and
confirm with
START/OK
key.
Show Setting
Technician
Set technician Code (Code 902), and confirm
with
START/OK
key.
Code
902
Lock Out
ON
Use one of
ARROW
keys to change from OFF
to ON. Confirm with
START/OK
key.
Adjust ?? & press OK
The default setting is “Lock Out -
NO”
.
Unlock in the same procedure.
“
2
” will be shown on the right side of the display, indicating that the
program is second level locked.
Level 3 – maximal rate locking
The operator can adjust the Maximal rate acceptable according to patient
needs:
Press
info
key twice.
Volume Infused
Show Setting
Use
ARROW
keys to scroll to "Adjust Setting"
option.
=> Adjust Setting
Enter "Adjust Setting" pressing
START/OK
key.
Buzzer level
=> More …
Use
ARROW
keys to scroll to "More".
Adjust Setting
Confirm selection with
START/OK
key.
Technician
Set technician Code (Code 901), using the
numeric keys, and confirm with
START/OK
key.
Code
901