VET-PRO
Operator’s Manual
48
Pressing
5
times
on the
info
key during
PCA
operation will display the time and date.
Time 8:30
Date 24.03.02
To clear the volume infused (VI): press the
info
key to display the volume infused (from stop or
setting mode) and then the
STOP/NO
key. The
screen will change to show volume infused zero.
Infused
Pressing the
STOP/NO
key after the program is completed , will erased the
VI. The option to erase the volume infused through info, is used in case the
last program was not completed.
F
Note:
Exit from the
info
mode is automatic after 10 sec, or can
be achieved by pressing the
info
key.
Program Delay
Delay of the program before the start is enabled in the Continuous
program (both in rate over volume and in volume over time), TPN and
the 25 steps program.
C
O
N
T
Volume 250ml
Press OK for Delay
To enable delay before start option, set delay
before start to on, as described on page 48
(Adjust Setting). After setting the program data
the pump will enable the delay setting.
C
O
N
T
Rate 25 ml/h
Volume 250ml
Time 10:00
Press OK for Delay
Time Delay 15:00
Press OK to start
Use the numeric keys to set the time delay (up
to 20 hours).
At this stage, if no value was set, pressing the
START/OK
key wil start the program
operation with no delay.
ml