There are two battery alarm conditions:
1.
The pump will warn that the battery is low a
minimum of 30 minutes before the “end
battery” alarm activates (when the battery
power reaches less than 7.1v).
Low Battery
Connect to charger
2.
After the ³ battery low´ warnings, when the
battery power reaches less than 6.05v, the
pump will cease operation and the "end
battery" alarm will be activated.
End Battery
Connect to charger
•
If the battery runs out,
End Battery
appears on the display. The pump stops
the infusion.
•
From the "end battery" state the user cannot restart the pump, until it has
been connected to mains.
G
Caution:
Leaving the battery in a discharged state for a long period of
time may damage the battery.
F
Note:
Whenever possible, use the pump connected to an AC power
supply via the charging unit. This preserves the battery power
supply for emergency use or for situations where the AC power
is not available.
F
Note:
Replace the battery once every four years.
Cleaning and Storage
Before the transfer of a pump to a new patient and periodically during use, clean
the pump by wiping over with a lint-free cloth lightly dampened with warm water
and a standard disinfectant / detergent solution.
I
Warning:
Before cleaning always switch off and disconnect from the AC
power supply.
G
CAUTION
Do not clean pump with chemicals such as Xylene, Acetone or
similar solvents. These chemicals cause damage to plastic
components and paint.