VET-PRO
Operator’s Manual
34
Volume 1256 ml
Max Rate: 108 ml/h
T
P
N
Press OK to start
5.
The pump will calculate all the set data and
display the Maximal rate for the program (to
be infused during Flat period). Press the
START/OK
key to start the program.
Setting Program Delay– Refer to page 48.
F
Note:
During the stop or data setting mode, use the
STOP/NO
key to
return one step backwards: First press will erase the last digit
displayed, once all the digits are erased pressing the
STOP/NO
key will display the previous screen .
Rate
U
P
2
ml/h
6.
During the TPN program operation, the pump
will display the actual rate and program
direction (Up/Flat/ Down).
Program operation will start from 2 ml/h and
will increase to the maximal rate.
Rate
. During the Flat session the rate will not change
and the pump will show the following message:
F
L
A
T
108
ml/h
Rate
.
During the last part of the program (T3), the
program will start to decrease the infusion rate
according to the set taper down time.
D
O
W
N
25
ml/h
F
Note:
T
P
N
Taper Down: 0:30
Press OK key
If the program must be interrupted
early, and the planned taper down
time must commence immediately
press the
STOP/NO
key continually
(2 sec) to stop the pump operation
and display Taper Down screen.
Set a new time or confirm the default setting. Press the
START/OK
key to start the taper down process immediately.