VET-PRO
Operator’s Manual
49
Delay 12:20
During the time delay, the
KVO
session will
run. Time left will count down, until 00:00. At
this stage the set program will automatically
start operation.
KVO 5 ml/h
Program Completed
End PROGRAM
When the program is completed, the following
message will appear on the display: press the
STOP/NO
key to mute the alarm.
Press STOP to mute
KVO 5 ml/h
The pump will automatically turn to a
KVO
mode when the program is completed. Press
the
STOP/NO
key to reset the pump and set a
new program.
Press STOP to renew
If the main program rate is higher than the KVO rate, set in the technician
mode, the rate will remain as set in the technician mode. If the main
program rate is lower than the KVO set rate in the technician mode, the
pump will operate at the program rate. KVO volume is limited to 5 ml.
F
Note:
Before operating the pump, make sure the IV bag contains 5
mls more than the VTBI– for the KVO (and delay time, if
applicable)
When the program is completed a special sound algorithm will be
performed: 4 single beeps followed by a 4 minute pause, then again 4
beeps, another 4 minute pause and a continuous alarm is activated.
When the audible alarm is activated, the display shows the “end program”
screen. Between the beeps and alarm pause the display shows the KVO
screen.
STOP
2 min
Press the
STOP/NO
key to stop the pump.
This ceases the pump operation and the
display shows STOP.
Press OK to start
Stop mode is limited to 2 min. After 2 min. an audiable alarm is sounded.
Press the STOP/NO key again, to mute the alarm.