VET-PRO
Operator’s Manual
56
Troubleshooting
Description
Result
Possible Cause
Remedy
Air in Line
Air in Line
Clean air
Air is present in
administration set
Disconnect patient,
press the
STOP/NO
key and remove the air
from Administration
set, as described on
page 17. After flushing,
press the START/OK
key to resume the
program.
Infusion
ceases
The roller clamp above the
pump is closed.
Open Roler clamp
Door Open
The door of the pump was
not closed prior to operation
Door Open
Close door
Infusion
ceases
The door has accidentally
opened during operation.
Close Door
Down stream occlusion
Roller clamp closed
Open the Roller clamp
Set kinked
Straight out the set
Down Occlusion
Press STOP to mute
Administration set loaded
incorrectly
Reload administration
set, as described on
page 17
Cannula blocked
Change Cannula
Infusion too cold, affecting
the elasticity of the tubing
Bring the tubing up to
room temperature
>19deg C. Warm the
liquid.
Infusion
ceases
**User may change
pressure setting
Adjust pressure setting,
as described on page 26
No key has been pressed
for 2 minutes, during
setting
Pump Unattended
Press start key
Pump
operation
has been
stopped
No key was pressed for 2
minutes.
Press the START/OK
key