22
On - In the on position, the emission current will track the beam cur-
rent at 200%.
Off - In the off position, the emission current will track the beam current
at 125%.
SW1-8:
On - In the on position, the front panel will remain locked following a
power cycle if it was locked with the serial communication command
NK1.
Off - In the off position. The front panel will be unlocked following a
power cycle.
All switches will be factory-set. The SW1-1 switch is for factory use only and
should never be changed.
Switch Settings for Switch Bank #2
SW2-1:
Not used.
SW2-2:
Controller Board Alarm Switch
On - In the on position, the controller board alarm will sound when a
Fatal error has been detected and will sound at power-up of the unit.
Off - In the off position, the controller board alarm will not sound when
a Fatal error has been detected or at unit power-up.
All switches will be factory-set. The SW1-1 switch is for factory use only
should never be changed.
Table 4.3: Factory Switch Settings For Switch Bank #1
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
Off
Off
Off
Off
On
Off
Off
Off
Table 4.4: Factory Switch Settings For Switch Bank #2
SW1
SW2
Off
On
NOTE
NOTE
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...
Page 95: ...91 3cm Source Hollow Cathode and Hollow Cathode Neutralizer...