Chapter 3: Installation
To avoid electrical shock, keep clear of “live” circuits. Follow all local lock-
out/tag-out procedures before installing any electrical devices.
Single Power Supply System Assembly
The MPS-3000 CE power supply is designed for rack mount. It is shipped
with the rack mount brackets attached.
1. Install the unit in the chosen location.
If cabinet installation is chosen, use the four holes in the rack mount
brackets to attach the unit to a cabinet.
Do not obstruct the cooling ports on the MPS-3000 CE power supply. An
insufficient supply of cooling air will result in the unit overheating, causing
a thermal shutdown and possible damage.
The power supply’s cooling air ports are located in the sides of the
chassis, and the cooling fan port is located in the rear panel of the
chassis.
2. Connect the following:
a. The source/beam cable to the source/beam connector (P33)
located on the unit’s back panel.
b. The neutralizer cable to the neutralizer connector (P32) located on
the unit’s back panel.
c. The interlock cable to the interlock connector (DB9) located on the
unit’s back panel.
d. The source/beam cable to the ion beam source (refer to
“Draw-
ings” on page 85
).
e. The neutralizer cable to the ion beam source (refer to
“Drawings”
on page 85
).
The shield in the source cable must be attached to facility ground with a
low impedance connection. If the cable is not attached correctly, the power
supply and facility may float at high voltages.
3. Connect the wires attached to pins 1 and 6 on the interlock cable to
the series-connected external interlock circuitry.
WARNING
CAUTION
WARNING
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...
Page 95: ...91 3cm Source Hollow Cathode and Hollow Cathode Neutralizer...