Command Name: MANUAL DATA LOG REQUEST
Syntax: RS<cr>
Units: N/A
Range: N/A
Response:
If the source and/or beam are on:
<lf><cr>^^^^^^^^TIME^^^^^^^CI^^^^^DI
^^^^^DV^^^^^BI
^^^^^BV^^^^^AV^^^^^AI^^^^^EI^^^^^
NI
<lf><cr>^^^^^^HH:MM:SS^^^^AMPS^^^AMPS^^
^VOLTS
^^^mA^^^^VOLTS^^VOLTS^^^mA^^^^^mA^^
^^AMPS
<cr><lf>^^^^^^XX:XX:XX^^^XX.XX^^^X.XX^^^XXX.X
^^XXXX^^^XXXX^^^XXXX^^^XXX^^^^XXXX^^XX.XX
<lf><cr><eot>
If the source and beam are off:
<lf><cr>^^^^^^^^^^^^^^^^^SOURCE^&^BE
AM^ARE^OFF^DDDD
<lf><cr><eot>
Where:
DDDD - Fatal Error Message. If there is no fatal error message when the RS
command is received, DDDD = blank spaces (32).
^ = ASCII BLANK SPACE (32)
Description: This command is used to request data logging parameters. It is
only valid when automatic data logging is disabled.
Command Name: RESET MPS-3000 CE
Syntax: RST<cr>
Units: N/A
Range: N/A
Description: This command is used to reset the power supply.