59
Command Name: MANUAL DATA LOG REQUEST WITHOUT HEADER
Syntax: RH<c>
Units: N/A
Range: N/A
Response:
With the source and/or beam on:
<cr><lf>^^^^^^AA:BB:CC^^^DD.DD^^^E.EE^^^F
FF.F^^^
GGG^^^HHHH^^^IIII^^^JJJ^^^^^KKK^^LL.LL
<lf><cr><eot>
Where:
With the source and beam off:
<lf><cr>^^^^^^^^^^^^^^^^^SOURCE^&^BE
AM^ARE^OFF^DDDD
<lf><cr><eot>
Where:
DDDD - Fatal error message. If there is no fatal error message when the RH
command is received, DDDD = Blank Space (32).
^ = ASCII BLANK SPACE (32)
Description: Used to request data logging parameters without header infor-
mation. If a value is off, it will be replaced with ASCII blank spaces (32).
Parameter
Model
Units
AA.BB.CC
Time since source was
turned on
Hours/
Minutes/
Seconds
DD.DD
Cathode Filament Cur-
rent
HC Heater Current
MPS-3000 CE FC
MPS-3000 CE HC
Amps
Amps
E.EE
Discharge Current
All
Amps
FFF.F
Discharge Voltage
All
mA
GGG
Beam Current
All
mA
HHHH
Beam Voltage
All
Volts
IIII
Accelerator Current
All
Volts
JJJ
Accelerator Current
All
mA
KKK
Emission Current
All
mA
ll.ll
Neut Filament Current
HCN Heater Current
PBN Filament Current
MPS-3000 CE FC
MPS-3000 CE HC
MPS-3000 CE PBN
Amps
Amps
Amps
^
ACSII Blank Space (32)
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...
Page 95: ...91 3cm Source Hollow Cathode and Hollow Cathode Neutralizer...