Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
193 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
Command
Remark
[SPM_ISRK]
Indicate whether the auto-return is completed after capture
[SPM_SNUM_xxxxxxxxx]
Omit the "-",
ex) 123-4567 => send [SPM_SNUM_1234567]
[SPM_SNU?]
Request the Serial Number
[SPM_SMRT_0000]
Sensor motor Run Torque Value Setting
[SPM_SMST_0000]
Sensor motor Stop Torque Value Setting
[SPM_CMRT_0000]
Collimator motor Run Torque Value Setting
[SPM_CMST_0000]
Collimator motor Stop Torque Value Setting
[SPM_RMRT_0000]
Rotator Run Torque Value Setting
[SPM_RMST_0000]
Rotator Stop Torque Value Setting
[SPM_HMRT_0000]
CEPH Sensor Run torque set
[SPM_HMST_0000]
CEPH Sensor Stop torque set
[SPM_UATT_xxxx]
Offset start position of the table on the CEPH sensor axis
[SPM_UAT?]
Read start offset of the table of CEPH sensor axis
[SPM_CSNM]
Normal resolution capture mode.
[SPM_ACSH]
Sharp arch
[SPM_ACST]
Standard arch.
[SPM_ACSQ]
Square arch
[SPM_ACCH]
Child’s arch
[SPM_TLA_]
TMJ lateral
[SPM_TPA_]
TMJ PA
[SPM_SLA_]
Sinus lateral
[SPM_SPA_]
Sinus PA
[SPM_CAAL]
Exposure all areas
[SPM_CAFT]
Exposure only the front part.
[SPM_CALT]
Exposure only the left side.
[SPM_CART]
Exposure only on the right side.
[SPM_ACMS]
Arct C mode
[SPM_ADMS]
Arct D mode
[SPM_LATF]
CEPH Lateral mode
[SPM_FRO_]
CEPH Frontal mode
[SPM_CAR_]
CEPH Carpus mode
[SPM_SMV_]
CEPH SMV mode
[SPM_CTB_]
Bright calibration starts in CT mode
[SPM_AXIS_xxxx]
Command for test purpose such as the motor movement of
each axis, and explained by sending [SPM_AXIS]
[SPM_CSEQ_xxxx]
Used to forcibly initialize and set a running sequence of
commands (command for equipment debugging).