Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
147 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
6. Save the bright cal data in a folder for each mode.
Norm Mode Path: C:\VCaptureSW\Acquisition\CEPH\WideCEPH\Norm\CAL
Fast mdoe Path: C:\VCaptureSW\Acquisition\CEPH\WideCEPH\Fast\CAL
7. When the normal bright cal acquisition is completed, click the
Check
button to create an auto
bad pixel map.
4. Click the Check button.
3. When the kVp and mA
values are confirmed, click
the Save button.
1. Check NumPoints=5
2. Check kVp, mA value.