Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
136 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
7.
Vertical Laser Beam Align
1. Run the Console Software.
2. Click the
Confirm
button and attach the ball phantom with a laser beam jig.
3. Click the laser beam mode to check whether the vertical beam is located at the reference line of
the ball phantom.
4. If the position is not correct, loosen the assembly point with an M4 wrench and move it left/right
so that it comes to the reference line.
5. If the angle of the vertical laser beam is not correct, loosen it with an M3 wrench as shown in
green in the picture below and adjust the angle.