Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
129 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
5.2. PANO Collimator Speed Alignment with EzAlign
1. After PANO Auto Alignment is completed, PANO Collimator Speed Alignment will be started
automatically.
2. When you see the message
HOLD THE BUTTON
as marked in the white box below, press and
hold the exposure switch until
RELEASE THE BUTTON
message appears.
3. If the system displays the
HOLD THE BUTTON
message again instead of the succeeding
message, repeat step 2.
4.
If the PANO Collimator Speed Alignment is completed successfully, the pop-up message
which notifies the completion of the alignment displays on the screen. Then, click the
OK
button to finish.
5. If the Error message appears, follow the guide. Then click the
Start
button again.