Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
69 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
8. Click the
Capture/Recon
tab and choose
#20: Bead(reconstruction.vxm)
for Recon mode.
9. Click the
Capture
button and make the exposure by pressing the X-ray exposure switch.
Stay outside of the X-ray shielding room to avoid unnecessary X-ray radiation
during the exposure.
10. Click the
PANO
button, and use the
Process
button to generate the average.raw of PROJ files.
Average after capturing the image
11. Click the
V Proj
button to see the averaged PROJ file.
Click the average image.
Check the Offset
(Standard range: 82 +-1 pixel)
Choose the
R.
Click the PANO button.
Click the Process button.