2.4. CEPH Ear-Rod Alignment
1. Run the Console Software.
2. Click
CEPH
>
High Resolution
>
Lateral
>
CONFIRM
and wait until
READY
button is enabled.
3. Turn CEPH Ear-Rod to the Lateral Mode position as shown in the figure.
4. Put a flat washer into the right Ear-Rod as shown below.
5. Click the
READY
button.
6. Perform the image acquisition according to the Console Software guidance.
7. When the image acquisition is completed, run the
VAKCAP.exe
file.
8. Click the
Panorama
tab >
V Pan
button to open the View16 program.
9. Click the
Brw
button and select the
Capture.raw
file to open the captured image.
•
Capture.raw
file path :
C:\VCaptureSW\Acquisition\Ceph\AnyCeph\Fast\PAN