Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
66 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
3.6.4. Bad Pixel Verification
•
If a BPM file is not created by clicking the
Check
button in the earlier
section, Bad Pixel Verification is not necessary.
1. Run
VAKCAP.exe
file. > Click the
Soft Calib
tab > Click the
V
button to open the
View16
program.
•
VAKCAP.exe file
Path :
C:\VCaptureSW\Acquisition\CT\WidePANO
folder
2. Choose
S
in the
M
box and click the
184x686BPM.raw
in the raw file list
3. Examine the sensor effective area to make sure that there is a bad pixel of a bad line.