Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
120 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
10. Click the 1 Point Bright Calibration data file on View16 and make sure that the Median value
meets the standard.
※
Bright standard value: 5 points with Cu 1 mm filter
Bright
Min
Max
Median
value
Cal Point 1
50
80
Cal Point 2
80
200
Cal Point 3
200
400
Cal Point 4
400
600
Cal Point 5
601 ~
If the median value does not meet the requirement, do the steps that follow:
Click the
Settings
tab.
Adjust
kVp0
value by 10 in the
[BrightCalibration]
field.
Click the
Save
button
.
Acquire Bright calibration data again.
11. Click the 5 Point Bright Calibration data file on View16 and make sure that the Median value
meets the standard.
If the median value does not meet the requirement, do the steps that follow:
Click the
Settings
tab.
Adjust
kVp1
value by 10 in the
[BrightCalibration]
field.
Click the
Save
button.
Acquire Bright calibration data again.
•
From the achieved PANO Sensor Bright Calibration data file, the 1
Point Bright Data file and 5 Point Bright data file should meet the
standard value. Another result from different points Bright Datafile
can be varied by the characteristics of each sensor.
max
ave
med
min
132
66.4
66
0