Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
116 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
3.4. PANO sensor: Dark calibration
1. Run the VACAP.exe file in the C:\VCaptureSW\Acquisition\PANO\WidePANO folder.
2. Click the
Settings
tab.
3. Click the
OK
and restart the program.
4. Click the
Capture/Recon
tab.