Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
158 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
※
Collimator Alignment standard value
Collimator
Min pixels
Max pixels
Threshold
90%
Left
1
5
Right
1
5
Upper
20
40
Lower
20
40
8. If the red pixel count is out of the permitted value, change the collimator position values by
[SPM_CAST_XXXXX].
Collimator Align can also be done using the EzAlign program provided by the
manufacturer.