Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
123 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
4.2. PANO Collimator Alignment with EzAlign
1. Select
PANO
and click the
Start
button: The system will move into its capturing position and
Collimator auto-align will be started.
2. Remove the Bite Block.
3. When you see the message
HOLD THE BUTTON
as marked in the white box below, press and
hold the exposure switch until
RELEASE THE BUTTON
message appears.
4. If the system displays the
HOLD THE BUTTON
message again instead of the succeeding
message, repeat step 3.
5. If the PANO Auto Alignment is completed successfully, the pop-up message which notifies the
completion of the alignment displays on the screen. Then, click the
OK
button to start
PANO
Collimator Speed Alignment
.
6. If the Error message appears, follow the guide. Then proceed the steps 1~5.