Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
122 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
※
Collimator Alignment standard value
Collimator
Min pixels
Max pixels
Threshold
70%
Left
2
2
Right
1
2
Top
1
10
Lower
1
10
If Collimator Alignment was performed and changed values, you must
perform Collimator Speed Calibration.
Collimator Align can also be done using the EzAlign program provided