Chapter 6. CEPH Image Evaluation
1.
Lateral Mode Start Position Evaluation
1.1. Lateral Mode Starting position standard
Parameter
LSL (Lower Specification Limit)
USL (Upper Specification Limit)
LAT start position
605
625
1.2. Phantom Specification
CEPH Phantom QUART Phantom
1.3. Test Method
1. Run Console Software and click the
CEPH
to go into ceph mode.
2. Set the High-Resolution/Lateral Mode/exposure condition 80 kVp, 10 mA, and click the
Confirm
button.
3. Place the CEPH Phantom on the ear rod.
4. Attach 0.8 mm Copper Filter on X-Ray Tube.
5. Click the
READY
button and acquire and save the image according to the Console Software
guidance.
6. Run View16.exe and open the capture.raw file to measure the center position of the Phantom.
1.4. Test Result
Parameter
LSL
Measurement value
USL
Pass/Fail
LAT start position
605
625
605~625 pixel