Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
190 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
2.
vatech A9(PHT-30CSS) Command-List
Command
Remark
[SPM_PANO]
Command to enter PANOrama mode
[SPM_CEPH]
Command to enter CEPH mode
[SPM_ARCT]
Command to enter Arch-CT mode
[SPM_RET_]
Move to the shooting start position for each mode
(PANOrama, CEPH, CT)
[SPM_ERDY]
Exposure switch press-ready command
[SPM_ERDC]
Exposure switch press cancel command
[SPM_PCON]
Command to let capture S/W run
[SPM_PCOF]
Command that the exposure software has ended (Cancel
switch press-ready cancel fu laser off function)
[SPM_HV___0000]
Tube Voltage Setting Command (ex) 0600 = 60kVp
[SPM_HA___0000]
Tube Current Setting Command (ex) 0400 = 4mA
[SPM_LON_]
Laser on depending on the capture mode
[SPM_LOF_]
Turn off all lasers
[SPM_CPON]
sensor power on (Operation for each mode)
[SPM_CPOF]
sensor power oFF (Operation for each mode)
[SPM_CSON]
CT sensor power on (Independent of the current mode)
[SPM_CSOF]
CT sensor power off (Independent of the current mode)
[SPM_PSON]
PANOrama sensor power on (Independent of the current
mode)
[SPM_PSOF]
PANOrama sensor power off (Independent of the current
mode)
[SPM_HSON]
CEPH sensor power on (Independent of the current mode)
[SPM_HSOF]
CEPH sensor power off (Independent of the current mode)
[SPM_PVER]
Know your motherboard firmware version (must be used
when the device is down. Transmission during operation may
cause equipment malfunction.)
[SPM_IVER]
Knowing the Firmware Version of the Inverter Board
[SPM_TVER]
Know the firmware version of the Main Board, Collimator
Board, Column Board, Inverter Board, MP3, Keypad
[SPM_TUTS]
Tube feedback test with exposing switch
[SPM_XTST_0000]
exposure time test. Input the command
[SPM_XTST_number] used for authentication
(Number is exposure time, 1 second = 0100, 10 seconds =
1000)
[SPM_XON_]
Turn on X-ray exposure
[SPM_XOF_]
Turn off X-ray exposure
[SPM_TMP?]
Tube temperature requirements