Model 990 dCLD II Component Leak Detector
5-6
5.3.1 Removing the Spectrometer Tube Assembly
The spectrometer tube operates at a very high vacuum produced by the high vacuum pump.
Service of the spectrometer tube requires that this vacuum be vented to the atmosphere.
NOTE
Rebuilt spectrometer tubes are available from Varian on an
exchange basis. Contact Varian Customer Service at
1-800-8VARIAN (1-800-882-7426) for details.
1. Loosen the hold-down screw
➀
to remove the spec tube block connector from the
spectrometer assembly (Figure 5-1).
Figure 5-1 Removing the Spec Tube Block Connector
2. Remove the turbo fan power connection and the turbo pump power connection from
the spec tube block connector.
➀