Model 990 dCLD II Component Leak Detector
1-7
Recommended
Ambient Operating
Conditions
Operating Temperature
Control Unit:
0 °C to 50 °C (32 °F to 122 °F)
Turbo Spec Tube Assy.:
5 °C to 35 °C (41 °F to
95 °F)
Optional Front Control Panel:
0 °C to 35 °C
(32 °F to 95 °F)
Maximum Relative Humidity
80% for temperatures up to 31 °C (88 °F),
decreasing linearly to 50% HR at 40 °C (104 °F).
No hoarfrost, dew, percolating water, rain, solar
radiation, etc.
Atmospheric Air Pressure
75 kPa to 106 kPa
Pollution Degree
II
Installation (Overvoltage)
Category II in accordance with UL3101-1
standard.
Heat Load
350 W
Power Requirements
Voltage Range
115 / 230 VAC
47 Hz to 63 Hz
Maximum Current
3 A/1.5 A, 350 Watts
Fuse Rating
5 A, 250 VAC, slo-blo
®
Conformance
Standards
Meets applicable UL, CSA, and CE Standards
Dimensions
Control Unit
19" rack, 5.25" high by 16.5" deep
48.26 cm rack, 13.33 cm by 41.91 cm
(including mating connectors)
Turbo Spec Tube Assembly
19" wide by 9" high by 8" deep
48.26 cm by 13.33 cm by 11.43 cm
Optional Front Control Panel
17.2" wide by 5.25" high by 4.5" deep
43.69 cm by 22.86 cm by 20.32 cm
(including mating connectors)
Weight
Control Unit Assembly
27 lbs. (12.25 kg)
Turbo Spec Tube Assembly
17 lbs. (7.71 kg)
Optional Front Control Panel
7 lbs. (3.18 kg)
Cables (turbo and valves)
14 lbs. (6.35 kg)
Summary of Contents for 990 dCLD II
Page 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 161: ......