Model 990 dCLD II Component Leak Detector
5-8
5. Remove the 2 slotted screws
➀
that mount the spectrometer tube magnet assembly to
the spectrometer tube body (Figure 5-4).
Figure 5-4 Slotted Screws that Attach the Magnet Assembly to Spectrometer
6. Loosen the tuning magnets
➀
by using a hex wrench to back out the set screws
¡
(located on both sides of the assembly) Figure 5-5. Do not remove the set screws and
tuning magnets completely.
Figure 5-5 Set Screws that Secure the Tuning Magnets
7. Carefully slide the magnet assembly off of the spectrometer tube body.
➀
➀
➁