Model 990 dCLD II Component Leak Detector
5-15
5.3.6 Examining and Cleaning the Spectrometer Parts
1. Using the Scotch-Brite™ pad included with the spectrometer tube cleaning kit, polish
away any discolored areas inside the spectrometer tube cavity.
2. Using the Scotch-Brite
pad, polish away any discolored areas of the ground slit plate
Figure 5-17 Discolored Ground Slit Plate
CAUTION
The ground slit plate is very thin. Take care not to bend or
disfigure it during cleaning.
3. Using the Scotch-Brite pad, polish away any discolored areas of the magnetic pole
Figure 5-18 Discolored Ground Magnet Pole Piece