Model 990 dCLD II Component Leak Detector
4-17
4.6.5.1 Bar Graph Display Set-Up
Touch the
LINEAR
BAR GRAPH DISPLAY
toggle box to switch the bar graph display
between LOG BAR GRAPH DISPLAY mode and LINEAR BAR GRAPH DISPLAY mode.
Note, this change is effective immediately upon selection.
4.6.5.2 Leak Rate Analog Output Voltage Selection
Touch the
LEAK RATE OUTPUT
selection
box located at the left of Output Control Set-Up
screen to switch the output voltage at the I/O port (J3 or J4) on the back of the control unit
among the selections listed below.
LINEAR ANALOG
OUTPUT
Touch the box to switch between LINEAR ANALOG OUTPUT or
LOG output voltage selections, which are 1V/DEC LOG, 2V/
DEC LOG and 3V/DEC LOG.
Convert the voltage to the leak rate using the appropriate chart in Section 3.7 “Analog Leak
Rate Output Voltage” on page 3-21.
4.6.6 Zeroing
Touch the
ZEROING SCREEN
box to display up the Zeroing screen shown in Figure 4-15.
Figure 4-15 Zeroing Screen
Touch the
AUTO-ZERO<0
box to switch AUTO-ZERO<0 between ACTIVE and INACTIVE
states. Touch
BACK
to return to the Main Menu. Note, this change is effective immediately
upon selection.
BACK
LEAK RATE
OUTPUT
CAL
LEAK
SET-UP
REJECT &
AUDIO SET
POINTS
RANGING
SET-UP
CONTROL
SET-UP
TO
SERVICE
MENU
ZEROING
SCREEN
UNITS
SET-UP
ZEROING
BACK
AUTO-ZERO
<Ø
ACTIVE
Summary of Contents for 990 dCLD II
Page 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 161: ......