Model 990 dCLD II Component Leak Detector
2-4
2.2.1 Single Mechanical Pump Configuration
Figure 2-1 Single Pump Configuration
Note: Foreline and Cal Leak valves are under the control of the 990 dCLD II.
Table 2-3 Single Mechanical Pump System Valve Sequence
State
Vent
Valve
Rough
Valve
Test
Valve
Remarks
Pre-Vented
Closed
Closed
Closed
The Pre-Vented state allows time to ensure that all valves
are closed before any open. The time spent in the
pre-vented state depends on valve actuation time.
Vented
Opened
Closed
Opened
The Vented state places the fixture at atmosphere. The
Test Valve keeps the turbo backed.
Pre-Roughing Closed
Closed
Closed
The Pre-Roughing state allows time to ensure that all
valves are closed before any open. The time spent in the
pre-roughing state depends on valve actuation time.
Roughing
Closed
Opened
Closed
The Roughing state brings the fixture to vacuum. The
turbo is not being pumped. The maximum time here is
approximately 3 minutes.
Test
Closed
Opened
Opened
The Test Valve re-opens at the appropriate foreline
pressure, backing the turbo and exposing the fixture to
the foreline.
Primary
Pump
Sp
e
c
Tu
rb
o
Cal Le
ak
Cal
Leak
Foreline
Turbo Spec Tube Assembly
Test
Valve
Rough Valve
Vent Valve
Fixture
Length is Proportional
Keep Distance as Short as Possible
(in box)
Tu
be
to Sensitivity
Summary of Contents for 990 dCLD II
Page 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 161: ......