Model 990 dCLD II Component Leak Detector
4-19
4.7.2 Manual Tuning
Touch the
MANUAL TUNING
box to view the Manual Tuning screen shown in Figure 4-18.
Figure 4-18 Manual Tuning Screen
You may need to manually tune the 990 dCLD II if:
❑
The control unit is connected to a spectrometer assembly in the field other than the
one with which it was tested
❑
It cannot successfully CALIBRATE
❑
The gain is greater than 5.0
❑
A board in the Platform is changed
❑
The ion source is changed
Refer to Section 5.6 “Manual Tuning” on page 5-20 for detailed instructions on Manual Tun-
ing.
Touch the FACTORY PRE-SET VALUES box to change the values to those set at the factory.
Note, this change is effective immediately upon selection.
The Manual Tuning screen allows you to change the REPELLER V, EMISSION I, ION V,
VAR FOCUS values.
BACK
CALIBRATE
MANUAL
GAUGE
CAL
MANUAL
TUNING
MANUAL TUNING
REPELLER V:
EMISSION I: X.XXX
ION V: XXX
VAR FOCUS: XXX
XXX
FACTORY
OK
DN
UP
ESC
FIL
1
ACTV
BACK
PRE-SET
VALUES
GAIN X.X