Model 990 dCLD II Component Leak Detector
A-8
The commands listed in Table A-4 are used to cause certain leak detection actions.
Table A-4 Leak Detection Actions
Command
Action
AUTO
Initiates Auto-Ranging mode. Success is indicated by the
ok
<CR><LF> response.
With PARALLEL ENABLE, the response is
cant
. AUTO is the factory default setting.
CALIBRATE
Performs an automatic calibration as specified by the parameters previously set.
Success is indicated by a
yes
<CR><LF> response to ?CALOK<CR>.
With PARALLEL ENABLE, the response is
cant
.
DECREMENT
Subtract 1 from the displayed leak rate exponent (making it more negative). Success
is indicated by the
ok
<CR><LF> response.
Has no effect in auto-ranging mode.
Has no effect upon reaching the most sensitive range.
With PARALLEL ENABLE, the response is
cant
.
EXTERNAL
Causes the control unit to use the external calibrated leak
FPEAK
Adjusts the variable focus ion source voltage to cause a maximum response to
helium. With PARALLEL ENABLE, the response is
cant
.
IDLE
Causes the control unit to power off the turbo pump and spectrometer tube and retain
operating parameters. ENABLE-IDLE must be set first.
INCREMENT
Add 1 to the displayed leak rate exponent (making it less negative). Success is
indicated by the
ok
<CR><LF> response.
Has no effect if in auto-ranging mode.
Has no effect upon reaching the least sensitive range.
With PARALLEL ENABLE, the response is
cant
.
INTERNAL
Causes the control unit to use the internal calibrated leak.
MANUAL
Terminates the Auto-Ranging mode. Success is indicated by the
ok
<CR><LF>
response.
With PARALLEL ENABLE, the response is
cant
.
RUN
Causes the control unit to start up the turbo pump and spectrometer tube from IDLE
mode. ENABLE-IDLE must be set first.
SYTCATM
Sets the current system thermocouple reading to represent atmospheric pressure.
Success is indicated by the
ok
<CR><LF> response.
If the response is
failed
<CR><LF>, replace the gauge.
SYTCZERO
Sets the current system thermocouple reading to represent a pressure that is too low
for a thermocouple to read. Success is indicated by the
ok
<CR><LF> response.
If the response is
failed
<CR><LF>, replace the gauge.
TUNE
Adjust ion source voltage to cause a maximum response to helium.
With PARALLEL ENABLE, the response is
cant
.
ZERO
Set the current leak rate measurement to be 0.0 atm cc/sec. Success is indicated by
the
ok
<CR><LF> response.
With PARALLEL ENABLE, the response is
cant
.
Summary of Contents for 990 dCLD II
Page 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 161: ......