Model 990 dCLD II Component Leak Detector
5-14
3. Remove the O-ring from the magnetic pole piece and place both the O-ring and the
magnetic pole piece on a clean surface (Figure 5-15).
Figure 5-15 Removing the O-ring from Magnetic Pole Piece
4. Repeat steps 1 through 3 to remove the second magnetic pole piece (Figure 5-16) from
the other side of the spectrometer tube body.
Figure 5-16 Removing the Second Magnetic Pole Piece
5. Remove the O-ring from the magnetic pole piece as in step 3.