Model 990 dCLD II Component Leak Detector
2-11
Figure 2-5 J3 Output Circuit Sketch
Table 2-8 J3 Outputs Summary Table
J3/
Pin
Output Name
Momentary
or Level
Description
2
REJECT #1
Level
Active when Leak Rate is over set point #1.
14
Ready
Level
Active when Fil and Turbo ok, i.e., the system is ready to test.
15
Zeroing
Level
Active when performing a ZERO function.
16
Leak Rate
Update
Valid data on BCD when low, becomes active during the time that
the BCD leak rate output is updated.
17
Filament Fault
Level
Active when selected Ion Source Filament is non-functional.
18
CAL OK
Level
Active when last calibration ok. It is set inactive at the start of a
CALIBRATE routine and at power on.
22
Calibrating
Level
Active when performing a CALIBRATE routine.
26
REJECT #2
Level
Active when Leak Rate is greater than set point #2.
27
+V Input
N/A
Customer-su5 to +24 VDC for J3.
28
+V Input
N/A
Customer-su5 to +24 VDC for J3.
29
REJECT #3
Level
Active when Leak Rate greater than set point #3.
30
REJECT #4
Level
Active when Leak Rate greater than set point #4.
33
HiVac OK
Level
Active when turbo is ok.
31
Analog Output
Analog
Analog Leak Rate output voltage.
32
Analog Return
Ground
Analog Leak Rate output voltage return.
Summary of Contents for 990 dCLD II
Page 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Page 161: ......