Model 990 dCLD II Component Leak Detector
2-7
2.3 Size of Mechanical Pumps
When selecting rough and forepumps for your system, the pumps listed in Table 2-6 are
recommended. These pumps can be paired (within type) in any combination for forepump
and rough pump except where noted. For more information, contact Varian Customer Ser-
vice 1-800-8VARIAN.
Table 2-6 Recommended Mechanical Pumps
Type/Model
Pump
Recommended for
Maximum
Pumping Speed
Remarks
Varian
Part Number
Fore
Rough
Rotary Vane
DS-102
l
100 L/min (3.6 cfm)
Do not pair with a Rough
Pump size
≤
DS-202
9499305
DS-202
l
l
200 L/min (7.0 cfm)
9499320
DS-302
l
l
300 L/min (10.6 cfm)
9499325
DS-402
l
400 L/min (14.1 cfm)
9499330
Dry Scroll
PTS 300
l
l
300 L/min (10.6 cfm)
PTS03001UNIV
PTS 600
l
l
600 L/min (21.2 cfm)
PTS06101UNIV