5.3.4 Removing the Preamplifier
CAUTION
The preamplifier is a static sensitive device. Wear a grounding
device while the preamplifier is being handled.
1. Remove the 4 slotted screws that hold the preamplifier header piece, repeating steps 1
and 2 in “Removing the Ion Source” .
2. Carefully remove the preamplifier
➀
from the spectrometer tube body (Figure 5-13) and
place it carefully on a clean, safe, non-magnetic surface.
Figure 5-13 Removing the Preamplifier
5.3.5 Removing the Magnetic Poles
1. Remove the 4 slotted screws
➀
to free the magnet pole piece
➁
Figure 5-14 Magnetic Pole Piece
2. Pick up the spectrometer tube body and tilt the side over to remove the magnetic pole
piece.
➀
➀
➁