A two-piece terminal block is used. Perform wiring with the plug pulled out from the option.
Use of the operating tool included is recommended to facilitate wiring.
1.
Strip the casing 6.5 to 7.5mm on the cable to
be connected.
Twist the bare wire to fit the terminal hole.
2.
Press the operating tool attached to the
operation slot at the upper part with a finger
and push down the spring.
3.
While pressing the operating tool, insert the
electric wire into the insertion opening until it
hits the wall.
4.
The wire will be connected once the operating
tool is released. For confirmation, tug the electric wire gently.
The wire can be connected likewise without the operating tool by using a tool such as a flathead
screwdriver to press down the spring from the top.
■
Zero scale value, full scale value
≪
Setting value
≫
-99999 to 99999 (where zero set value < full scale set value)
Zero scale value:
Set the indicated value for the time of output of 0V.
full scale value:
Set the indicated value for the time of output of 10V.
* Set torque, rotation speed and power, respectively.
How to set
CAUTION
- Cable can be from 28 to 14AWG (0.08 to 1.5mm
2
).
- It is not necessary to solder the cable wires or to fix a solderless terminal.
- If several cables to be inserted to the same hole, twist those cable wires together
and insert.
- Connect the cable with the power supply of the main body turned off without fail.
- For the protection from the damage, do not operate the lever without removing the
plug.
6.5 to 7.5mm
SET.
Option Setting (D/A Output)
Page 1
→
System Setting
→
→