About calibration
"Calibration" is performed for matching the TM700 to a sensor.
Check the UTM
Ⅱ
to be used.
●
Equivalent input calibration
In this calibration procedure, only the rated capacity value (one to be displayed) of a sensor is
input using the keys.
For example,
for UTM
Ⅱ
-0.1Nm:
0.100Nm
100.0mNm
for UTM
Ⅱ
-100Nm: 100.0Nm
0.100kNm
The valid number of digits is 5, and the number of digits after the decimal place is up to 3.
For example, display settings such as
1000.00mNm (6 digits) and
0.1000Nm (4 digits after the decimal place)
are not allowed.
4-1.
Calibration Procedure of the Torque
Release the calibration protection.
Set the calibration value No.
(Set "0" for use with only one selection.)
(optional if there is no change)
Set the unit of the values displayed.
Register the capacity of the sensor.
At this time, also input the decimal place.
Set the zero point of the sensor in no-load
condition (with the sensor unloaded).
Set the minimum unit of torque such as scale interval
and scale division. (optional if there is no change)
The load limit to which digital zero is enabled from the
zero calibration point is set. (optional if there is no
change)
4
CALIBRATION
*
…It is indispensable.
③
Unit
*
④
Equivalent input
*
⑤
Zero calibration
*
⑥
Increment
⑦
Digital zero limit
calibration
①
Calibration protect
②
Calibration selection