5-9.
Comparison Functions
Set the HI/LO limit of torque and rotation speed. (Hold synchronized)
≪
Setting value
≫
- Torque
HI limit/LO limit:
-99999 to 99999
- Rotation speed
HI limit/LO limit:
0 to 99999
<
HI/OK/LO output conditions
>
- HI
Indicated value
>
HI limit
- LO
Indicated value
<
LO limit
- OK
LO limit
≦
Indicated value
≦
HI limit
Overload of the torque and rotation speed is monitored. (Hold not synchronized)
When they exceed the alarm HI limit or fall below the alarm LO limit, the external output "ALM"
turns ON.
≪
Setting value
≫
- Torque
Alarm HI limit:
-99999 to 99999 (invalid at 99999 (initial value))
Alarm LO limit:
-99999 to 99999 (invalid at -99999 (initial value))
- Rotation speed
Alarm HI limit:
0 to 99999 (invalid at 99999 (initial value))
Alarm LO limit:
0 to 99999 (invalid at 0 (initial value))
- For initial value, since both the alarm HI and LO limits are invalid as default settings,
set them again for use.
- Since the alarm HI limit and LO limit are independent, simple use of either one of them
is also possible.
■
HI limit, LO limit, alarm HI limit, alarm LO limit
How to set
With ±LOAD, OFL1, or OFL2 on torque, torque HI, LO is forcedly turned on.
With OFL2 on rotation speed, rotation speed HI, LO is forcedly turned on.