Single: Plots a graph only once without using the trigger.
Starts take-in by start/stop key input or with START/STOP external input turned ON.
Take-in will also be completed after plotting a graph for once.
Single Trigger: Plots a graph only once using the trigger.
Starts take-in by start/stop key input or with START/STOP external input turned ON. Compares
the trigger level and torque, and starts plotting a graph when the conditions for trigger slope are
satisfied.
Take-in will also be completed after plotting a graph for once.
Example) Graph trigger slope
→
Falling edge
■
Interval time
If you select "Continued" or "Continued trigger" in the graph function setting, set the graph plotting
operation interrupting time from clearing the screen until moving to the next graph plotting
operation. During this time, the graph screen is held.
≪
Setting value
≫
0.1 to 99.9 [sec.]
How to set
t
+
(Key)
Plotting section
Plotting time
OFF
ON
OFF
ON
1.0msec more
START/STOP
(External input)
START/STOP
t
+
START/STOP
(External input)
Plotting section
X(TM) end point
Trigger Level
Trigger wait
1.0msec more
Screen hold, Trigger wait
SET.
Graph Setting
Page 1
→
→
Meas. Setting
→