19
4
CALIBRATION
19
C
A
LI
B
R
A
TI
O
N
Chapter
4
4-4.
Unit
Set the unit for performing calibration.
After this setting, make sure to perform calibration.
≪
Setting value
≫
mNm, Ncm, Nm, kNm, kgm, kgcm, gcm
How to set
1.
Press the [SET.] button. (see P.17)
2.
Press the [Calibration] button.
3.
Press the [Unit] button.
4.
Select the unit and determine with the [OK]
button.
SET.
Calibration
Page 2
→
→
* Even if the unit is changed,
the display value (calibration value)
will not change.
Power is calculated by performing internal conversion based on the unit settings.
Therefore, do not change these settings improperly after calibration. If they are
changed, perform calibration again.
For unit conversion, see "4-12.Unit Conversion Table" on page 27.
Point
Summary of Contents for TM700