Chapter
5
■
Graph mode
Set the graph plotting mode.
≪
Setting value
≫
Continued, Continued Trigger, Single, Single Trigger
How to set
Continued: Plots a graph continuously without using the trigger.
Starts take-in by start/stop key input or with START/STOP external input turned ON. When
plotting is complete, the screen will be cleared after an interval time and the next plotting will
start. Stops take-in by start/stop key input or with START/STOP external input turned OFF.
Continued Trigger: Plots a graph continuously using the trigger.
Starts take-in by start/stop key input or with START/STOP external input turned ON. Compares
the trigger level and torque, and starts plotting a graph when the conditions for trigger slope are
satisfied. When plotting is complete, it will be on trigger standby again after an interval time.
Stops take-in by start/stop key input or with START/STOP external input turned OFF.
Example) Graph trigger slope
→
Rising edge
SET.
Graph Setting
Page 1
→
→
Meas. Setting
→
t
+
START/STOP
START/STOP
Start
Plotting section
Plotting section
X(TM) end point
X(TM) end point
Screen clear
Interval time
(Key)
(External input)
OFF
ON
OFF
ON
Take-in complete
(Screen hold)
+
Plotting section
Plotting section
Interval time
Trigger Level
t
START/STOP
START/STOP
(Key)
(External input)
OFF
ON
OFF
ON
(Screen hold)
Take-in complete
Trigger wait