Blood Pressure Monitor

Instruction Manual

iProvèn BPM-337

Manual Ver.: 2.3

Issuing Date: 2017/10/14

©2017. All rights reserved.

Thank you for making iProvèn your top choice. Please read the 
instructions carefully in order to accurately and safely utilize this 
equipment.

The BPM-337 measures the blood pressure and the heart rate 
and saves the results. The BPM-337 off ers real time and accurate 
readings that are as convenient as possible thanks to the 
Oscillometric Measuring method. Blood pressure measurements 
may not be valid or accurate if they are not performed in 
accordance with the instructions provided in this manual. Please 
keep these instructions handy for future reference. You can also 
fi nd the digital version of this manual on www.iproven.com

It is our passion to develop high quality products for home use. 
Our products are manufactured at the highest technical standards 
of professional quality, durability, and consistency. They are also 
designed with elegant simplicity in mind, making them easy to 
use at home.

To help you get the most from our products, we provide clear 
instructions with each device. The manual also includes helpful 
information that contributes to your overall health awareness.

In order to make sure that our products are tailored to your needs, 
we welcome your feedback. If you have any issues, questions or 
recommendations, please share your thoughts with us at 
www.iproven.com

iProvèn - Professional Care Brought Home

iProvèn is a Masena Invest Brand

Intended use of the BPM-337 blood pressure  

monitor

The BPM-337 is a digital monitor that is used for measuring blood 
pressure and heart rate by cuffi  ng the device around wrists of 
diff erent sizes from 5   ˝-8½˝  (13.5cm to 21.5 cm).  The BPM-337 is 
intended for home use and for adults only. 

How this device works

The product uses the Oscillometric Measuring method to detect 
blood pressure. When the measurement is started, it calibrates to 
“zero pressure” which is the natural air pressure. As soon as it starts 
infl ating the wrist cuff , the device detects the pressure oscillations 
according to the beat-to-beat pulsatile and so it determines the 
systolic, diastolic pressure and the heart rate. 

Irregular Heart Beat detection

The device is equipped with irregular heart beat detection, IHB. 
The algorithm of the device compares the longest and shortest 
intervals of registered pulse waves (the time interval) and 
calculates the standard deviation. If the diff erences in the time 
intervals are more than 25% you have an irregular heartbeat 
and the IHB sign will appear in the display. If the device detects 
irregular heartbeat during consecutive measurements and you are 
following the correct procedure, please consult your doctor.

Setting Up the Device

Firstly, make sure you set the time and data correctly so the device 
can register the results with the correct time.

1.  With the monitor OFF, press and hold the SET button for 3 

seconds. It will enter the SETTINGS Mode.

2.  Press or hold the MEM button to adjust the Hour digits that are 

blinking on the display.  Press SET to confi rm. 

3.  Repeat step 2 for setting the Minutes;  Month; Day and Year.
4.  Press SET one more time to select the desired measurement unit. 

Probably “mmhg”.

5.  Once the Year is set, the display will show “mmHg”.  This is the 

standard measurement unit. Press SET to save. Or press MEM to 
change to “kPa” and then press SET to save.

6.  Once the measurement unit is set, the LCD will show  “DONE” 

and the device will shut off .

Preparing the measurement

Before tying the cuff  to your wrist, make sure you remove all 
accessories from the wrist (bracelets, watch, etc.). Use the device
on the wrist that doesn’t have poor circulation.

1.  Roll up the sleeve to reveal the skin of the wrist.
2.  Make sure to tie the cuff  to 0,4”-0,6” (1-1.5cm) from

your wrist joints.

3.  Tie the cuff  to your wrist and keep the palm facing up.
4.  Tie the cuff  to your wrist tightly but not painfully.
5.  Place your elbow on a table and make sure that the wrist is

at the same level with your heart. 

6.  Rest 3 minutes; keep your palm upwards, back straight and 

breathe deeply for 5 or 6 times before starting the measurement.

Taking a measurement 

1.  After resting press the Start Button and the measuring 

procedure will start.

2.  First the measured air pressure will calibrate to zero and then 

the device will infl ate and start reading the blood pressure and 
heart rate.

3.  When the measurement is completed the results will appear in 

the screen. The measurement is also saved in the memory.

Tips and Timing

•  Make sure you do not move or speak and that you are in a 
        silent room during a measurement.
•  Do not cross your legs during the measurement.
•  Do not use the device in a toilet or while doing any sort of 
        activity including talking or moving. 
•  Do not use the device in a cold environment. 
•  To get insight in the trend of your blood pressure, it is 
        best to measure your blood pressure at about the same time(s)  
        every day.
•  For accurate results, avoid using the device within 1 Hour
        after eating. 
•  You shouldn’t measure immediately after drinking tea or coff ee
        and after smoking. 

Extra notes on the device

•  The unit should not be used for prolonged monitoring during
      medical emergencies or operations.
 •     The cuff  being infl ated for a prolonged time will lead to numb  
       ness of wrist and fi ngers, causing anesthesia, pain and 
       ecchymosis.
•      Use the device according to the instructions of this manual to 
       guarantee an effi  cient performance and durability of the device.
•       The cuff  complies with the requirements of ISO 10993-5:2009 
       and ISO 10993-10:2010.
•       The cuff  does not cause any potential allergic reaction or 
       contact injury.
•       There is no need for calibration during the two years of ‘
        guaranteed service.

Components of the BPM-337

The Monitor Components

The LCD Display

The Cuff  Components

How to Install and Replace 

Batteries:

 1. Open the battery cover
2.  Insert the two AAA batteries conform the right polarity
3.  Close the battery cover

          

CAUTIONS

•  Adult-use ONLY;
•  Non-invasive measuring and monitoring of arterial blood
      pressure;
•  Use only on the WRIST!
•  Consult your doctor about the use of this device;
•  When the pressure on the cuff  exceeds 40 kPa (300 mmHg), 
      the unit will automatically defl ate;
•  If the unit doesn’t defl ate when exceeding the 40 kPa (300 mmHg),
      detach the cuff  from the wrist and press the START/STOP button   
      to stop infl ation;
•  Do not use the monitor under conditions of strong 
      electromagnetic fi elds;
•   Electromagnetic fi elds will interfere with the signal and
       electrically fast transient/burst signal;
•   The maximum temperature that the applied part can achieve is  
       42.5 ºC while the environmental temperature is 40 ºC;
•   The device is not AP/APG equipment;
•   It is not suitable for use in the presence of a fl ammable anesthetic 
       mixture with air (or oxygen, nitrous oxide);
•  Keep the Unit OUT OF REACH of infants, children or pets since
      inhalation or swallowing of small parts is dangerous or even fatal;
•  Use only accessories or detachable parts specifi ed and authorized 
      by the Manufacturer;
•  This Device should NOT be used on females suspected of 
      pregnancy or pregnant! It may provide inaccurate readings and   
      can have eff ects on the fetus. 
•  If you are suff ering from arrhythmias such as atrial or ventricular 
      premature heartbeats or atrial fi brillation the measurement results   
      can be incoherent. In this case please consult your physician about 
      the results. 

CUFF

(Type BF applied part)

BATTERY

COMPARTMENT

Component List:

1. PCBA;

2. Air Pipe;

3. Pump;

4. Valve;  
5. Cuff. 

•  Avoid measuring after 30 minutes of taking a shower or bath.
•  For an accurate and clear result, please measure the same wrist 

every time. Blood circulation may diff er from wrist to wrist so you 
can get a diff erent result when you change wrists. 

Using the memory function

•  Make sure that the date and time is set correctly. 
•  Every measurement is automatically saved in the memory,
        with a maximum capacity of 60 measurements; 
•  To view the recorded data press the MEM button while the
       monitor is inactive. If the monitor is not yet inactive, fi rst
       deactivate it by pressing the start/stop button.
•  Use the buttons MEM for Up and Set for Down to read through 
       the records.

Deleting the records

In case you want to delete your recorded measurements, please 
follow these steps:

1.  Make sure the monitor is inactive; press START/STOP if needed.
2.  Press MEM to go to Memory.
3.  Press and hold the MEM button for 3 seconds. 
4.  “DEL ALL” will start blinking. Press SET to confi rm. 
5.  “DEL DONE” will show, meaning the stored data is deleted. The 

device will shut off . 

Keeping the device safe

Please make sure to place the device away from the sun. Store it in 
a dry place. When you want to clean the device, you should use a 
dry cloth. Do not place it in water or clean it with wet cloths.  Also, be 
careful not to shake or throw the device. For a better performance, 
keep it in a room with stable temperature and away from dust. The 
cuff  should not be cleaned as it may aff ect the accuracy of the reading.

High blood pressure

Low blood pressure

Measurement unit of blood pressure

(1mmHg=0.133kPa)

Hour:Minute (Month/Day/Year)

SYMBOL DESCRIPTION  EXPLANATION 

Pulse in beats per minute

Blood pressure monitor is detecting an 

irregular heartbeat during measurement.

Low battery and please replace the batteries.

Indicating the memory mode and the 

number of the measurement

Indicating the AHA category of the measurement

See Error Messages and FAQ

High blood pressure level

Low blood pressure level

Pulse in beats per minute

Low battery, please replace the batteries

Indicating the AHA category of the

measurement

See Error Messages and FAQ section

Blood pressure monitor is detecting an

irregular heartbeat during measurement

Indicating the memory mode and the 

number of the measurement

Hour: Minute (Month/Day/Year)

Measurement unit of blood pressure

(1mmHg=0.133kPa)

Measurement units

Memory

Error

IHB Detector

AHA indicator

Low battery

Pulse

Diastolic

Blood Pressure

Systolic

Blood Pressure

Time