SonixTablet Service Manual
00.053.129, Revision C
Appendix B: Electromagnetic Immunity Tables
B-1
APPENDIX B: ELECTROMAGNETIC IMMUNITY TABLES
Table B-1: EN 60601-1-2:2001 (Table 201)
Table B-2: EN 60601-1-2:2001 (Table 202)
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
SonixTABLET Ultrasound Systems are intended for use in the electromagnetic environment specified below. The customer or the user of the
SonixTABLET Ultrasound System should ensure that it is used in such an environment.
Emissions Test
Compliance Level
Electromagnetic Environment-Guidance
RF emissions CISPR 11
Group 1
The SonixTABLET Ultrasound System uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
Class A
The SonixTABLET Ultrasound System is suitable for use in all establishments other
than domestic and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
SonixTABLET Ultrasound Systems are intended for use in the electromagnetic environment specified below. The customer or the user of the
SonixTABLET Ultrasound System should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment-Guidance
Electrostatic
discharge (ESD)
IEC 601000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1kV for input/output lines
±2 kV for power supply lines
±1kV for input/output lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
±1kV differential mode
±2kV common mode
±1kV differential mode
±2kV common mode
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage DIPS,
short interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% UT (>95% dip in UT)
for 0,5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycle
<5% UT (>95% dip in UT)
for 5 sec
<5% UT (>95% dip in UT)
for 0,5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycle
<5% UT (>95% dip in UT)
for 5 sec
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the SonixTABLET
Ultrasound System requires continued operation during
power mains interruptions, it is recommended that the
SonixTABLET be powered from an uninterruptible power
supply (UPS) or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
environment.
Note:
U
T
is the AC mains voltage prior to application of the test level.
Summary of Contents for SonixTablet
Page 1: ...SonixTablet Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Page 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Page 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Page 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Page 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Page 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Page 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Page 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Page 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...