Chapter 8: DICOM
00.053.129, Revision C
SonixTablet Service Manual
8-6
The
DICOM Global Storage Settings
dialog specifies global image storage parameters.
Figure 8-5: DICOM Storage Settings – Global Storage Settings
Table 8-4: DICOM Storage Settings – Global Storage Settings
S
torag
e
Opt
io
n
s Show Error
Balloons
Select to enable the display of
DICOM Storage
error messages (e.g.,
Failed to connect to DICOM
).
Schedule
Select an auto-transfer setting:
End of Exam
,
Immediate
,
On Idle
.
Storage Folder
Select the location (local or remote) where the images will be stored.
Note:
If a value is specified, the
AE Configuration
and
Storage Commitment
dialogs are disabled
– images can not be stored to an
SCP
.
Insert (Symbol)
Enables the insertion of text symbol(s) not available on the keyboard (e.g., punctuation, symbols and
letters from other languages).
Summary of Contents for SonixTablet
Page 1: ...SonixTablet Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Page 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Page 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Page 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Page 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Page 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Page 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Page 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Page 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...