SonixTablet Service Manual
00.053.129, Revision C
Chapter 12: Troubleshooting Issues
12-3
12.7 IMAGE IS GRAINY OR MISSING
Check to ensure
Clarity
is set to
Medium
or
High
:
•
A.4.1 Adjust the Clarity Setting
.
Check to ensure the
Clarity License
is
Enabled
:
•
A.4.2 Check the Clarity License Status
.
Re-import
license.txt
to ensure that
Licenses
are not set to
Disabled
:
•
.
Check to ensure the transducer is not defective:
• test the transducer on another system and compare images
• submit test images for diagnosis to Ultrasonix Technical Support.
12.8 IMAGE HAS BLACK OR DARK AREAS
Black or dark areas on an ultrasound image are usually caused by dead elements in or delamination of the transducer. The
problem may be caused by either the transducer or the transducer MUX board, so both may have to be tested.
•
•
7.4.2 Testing the Transducer Board
Figure 12-1: Dead Element or Delamination Example
Summary of Contents for SonixTablet
Page 1: ...SonixTablet Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Page 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Page 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Page 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Page 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Page 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Page 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Page 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Page 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...